Metrohm 758 KFD Titrino User Manual
Page 130
3.2 Remote control commands
758 KFD Titrino
126
3.2.2.133. Setup.Mode.StartWait ON, OFF
Setup.Mode.FinWait ON, OFF
Holding points in the method sequence. If they are "ON", the sequence stops
until "OFF" is sent. Switching the instrument on sets both nodes to OFF:
.StartWait: Holding point right after starting a method or submethod in TIP
(holding point after AutoInfo !".T.GC").
.FinWait: Holding point at the end a method or submethod in TIP (holding
point after AutoInfo !".T.F").
3.2.2.134. Setup.SendMeas.SendStatus ON, OFF
Setup.SendMeas.Interval 0.08...4...16200, MPList
.SendStatus: ON means the automatic transmission of measured values (see
3.2.2.136 and 3.2.2.137) in the inputted interval is active.
.Interval: Time interval (in s) for the automatic transmission of associated
measured values defined under points 3.2.2.136 and 3.2.2.137.
The inputted value is rounded off to a multiple of 0.08. The small-
est possible time interval depends on the number of measured
values which have to be sent, on the baud rate, on the load on the
interface and on the type of device connection. With "MPList" the
measured values are sent at the time of their entry into the meas-
ured point list.
The automatic transmission is switched on/off with 'SendStatus'.
3.2.2.135. Setup.SendMeas.Select Assembly,
Titrator
Selection of the unit of which the measured values should be sent (3.2.2.136
and 3.2.2.137).
3.2.2.136. Setup.SendMeas.Assembly.CyclNo ON, OFF
Setup.SendMeas.Assembly.V ON, OFF
Setup.SendMeas.Assembly.Meas ON, OFF
Selection of the values from Assembly for the output in the set time interval
(see 3.2.2.134):
.CyclNo: Cycle number of the potential measurement. Together with the
cycle time (3.2.2.118), a time frame can be set up.
The cycle number is set to 0 on switching on the instrument and it
is always incremented as long as the instrument remains
switched on.
.V: Volume
.Meas:
Measured value associated to the cycle number.
The unit "assembly" must be preset (see 3.2.2.135).
3.2.2.137. Setup.SendMeas.Titrator.CyclNo ON, OFF
Setup.SendMeas.Titrator.V ON, OFF
Setup.SendMeas.Titrator.Meas ON, OFF
Setup.SendMeas.Titrator.dVdt ON, OFF
Setup.SendMeas.Titrator.dMeasdt ON, OFF
Setup.SendMeas.Titrator.dMeasdV ON, OFF
Setup.SendMeas.Titrator.T ON, OFF
Selection of the values from the titrator which are sent in the set time interval
(see 3.2.2.134, formats see 3.2.2.113):