3.2 Remote control commands
758 KFD Titrino
117
Reports which are sent from the Titrino are marked with space (ASCII 32) and
' at the beginning. Then an individual identifier for each report follows. Reports
which are triggered by RS232 ($G) have the same introducer but without pre-
ceding space, i.e. they start with '.
3.2.2.102. Info.CalibrationData $G
Info.CalibrationData.Inp1.pHas –20.00...7.00...+20.00
Info.CalibrationData.Inp1.Slope –9.999...1.000...+9.999
Info.CalibrationData.Inp1.Temp –170.0...25.0...+500.0
Info.CalibrationData.Inp1.Date read
only
Info.CalibrationData.Inp1.ElectrodeId read
only
identical for .Inp2 and .Diff
pH calibration data for measuring input 1. After the calibration, the data are
entered automatically together with the date of the calibration and the elec-
trode identification, see 3.2.2.42.
Calibration data can be entered. They are accepted with &Info.CalibrationData
$G. If calibration data are entered, the calibration date is deleted.
3.2.2.103. Info.PrepData.D0.Date read
only
Info.PrepData.D0.Time read
only
Info.PrepData.D1.Type read
only
Info.PrepData.D1.Date read
only
Info.PrepData.D1.Time read
only
identical to .D2
Data from the last correctly executed preparation.
3.2.2.104. Info.Checksums $G
Info.Checksums.MPList read
only
Info.Checksums.ActualMethod read
only
The checksums can be used to identify the content of a file unequivocally,
e.g. files with identical content
have identical results of the checksums. An empty file has checksum "0". The
calculation of the checksums is triggered with $G.
.MPList: Result of the checksum of the current measuring point list.
.ActualMethod: Result of the checksum of the current method in the working
memory. Identical methods with different method names have the
same results of the checksum.
3.2.2.105. Info.DetermData $G
Info.DetermData.Write ON,
OFF
Info.DetermData.MPList.1.Attribute read
only/read+write
Info.DetermData.ExV read
only/read+write
Info.DetermData.MPList.1.X read
only/read+write
Info.DetermData.MPList.1.Y read
only/read+write
Info.DetermData.MPList.1.Z1 read
only/read+write
Info.DetermData.MPList.1.Z2 read
only/read+write
for every measuring point
Determination data in hexadecimal format. A measuring point list is available
in mode SET, KFT and MEAS.