3.2 Remote control commands
758 KFD Titrino
106
3.2.2.36. Mode.Parameter.Measuring.Temp –170.0...25.0...500.0
Parameters for MEAS: Measurement temperature in °C. If a Pt100 or Pt1000
is connected, the temperature is measured.
The temperature is used for the temperature correction in pH measurements.
3.2.2.37. Mode.Parameter.Measuring.TDelta 1...2...999999
Parameters for MEAS: Time interval in s for the acquisition of a measured
value into the measuring point list.
3.2.2.38. Mode.Parameter.Calibration.MeasInput
1, 2, diff.
Parameters for CAL: Selection of the measuring input. "diff." means differen-
tial amplifier, see page 167.
3.2.2.39. Mode.Parameter.Calibration.CalTemp –20.0...25.0...120.0
Parameters for CAL: Calibration temperature in °C. If a Pt 100 or Pt1000 is
connected, the temperature is measured.
3.2.2.40. Mode.Parameter.Calibration.Buffer.1.Value 0...7.00...±20.00
Mode.Parameter.Calibration.Buffer.2.Value 0...4.00...±20.00, OFF
etc. up to 9 buffers
Parameters for CAL: pH of buffers. The first buffer which is set to "OFF" de-
termines the number of buffers in the calibration.
3.2.2.41. Mode.Parameter.Calibration.SignalDrift 0.5...2...999, OFF
Mode.Parameter.Calibration.EquTime 0...110...9999, OFF
Parameters for CAL: Criteria for measured value acquisition. Measured value
drift in mV/min, equilibration time in s. OFF means that the corresponding cri-
terion is switched off. If both criterions are on OFF, the measured value is ac-
quired immediately.
If the equilibration time has never been edited, it is automatically calculated by
the instrument to match the drift, see page 36. After it has been edited once, it
remains in force with the set value.
3.2.2.42. Mode.Parameter.Calibration.ElectrodeId
up to 8 ASCII char
Parameters for CAL: Electrode identification. It is classified under calibration
data, see 3.2.2.102.
3.2.2.43. Mode.Parameter.Calibration.SmplChanger ON, OFF
Parameters for CAL: Calibration at Titrino.
With "ON", there are no hold points in the calibration sequence for entries, the
first buffer is measured directly.