3.2 Remote control commands
758 KFD Titrino
129
.Right:
Relative specification of the width of the output medium (e.g. pa-
per width) for the length of the measured value axis. 1 means the
measured value axis is plotted over the entire width of the paper
(largest possible width). In extreme cases, the writing of the right
tick may lie outside.
.Feed:
Length of the time axis. Depending on the printer, the measure in
cm may not always be correct.
Curve
length
0.01 100
cm
0.1 10
cm
0.5 2
cm
1 1
cm
3.2.2.140. Setup.PowerOn $G
Simulation of 'power on'. The device has the same status as after power on:
The cylinder is filled, error messages deleted and the current sample number
set to 0. The method last used is ready for operation.
3.2.2.141. Setup.Initialise $G
Setup.Initialise.Select ActMeth, Silo, Calib, Config,
Assembly, Setup, All
Setting of default values for the following areas:
ActMeth: Current method. Parameters, calculations, and assignments for
the data output, operands C01...C19.
Silo:
The silo memory is deleted. Same function as delete entire silo.
Calib:
pH calibration data for all measuring inputs.
Config:
All values under &Config.
Assembly: All values under &Assembly.
Setup:
All values under &Setup.
All:
Values of the entire tree (except silo and method memory).
The action must be triggered with &Setup.Initalise $G.
3.2.2.142. Setup.RamInit $G
Initializes instrument, see page 153. All parameters are set to their default
value and error messages are cleared. The user and silo memories will be de-
leted. The user memory contains the default user methods from Metrohm.
3.2.2.143. Setup.InstrNo $G
Setup.InstrNo.Value serial
number, 8 ASCII characters
Instrument identification for report output.
Set the value with &Setup.InstrNo $G .
3.2.2.144. Diagnose.Report $G
Output of the report containing the adjustment parameters. The Titrino has to
be in its inactive basic state.