5.5. Continuous dosing with two Dosimats

765 Dosimat

58

Example:

Continuous dosing of 55 mL. Exchange units with volumes of 20 mL and 10 mL.

Key:

End volume

V = 55 mL

Exchange unit 1

V

1

= 20 mL

Exchange unit 2

V

2

= 10 mL

55

1. N = Int———— = 1

20+10

2. V

R

= 55 - 1*(20+10) = 25

3. 25 (V

R

) is larger than 20 (V

1

), i.e. Dosimat 2 is the final burette.

4. V-LIM for Dosimat 2 is: 55-20*(1+1) = 15

Settings:

Dosimat 1 (start Dosimat)

Dosimat 2 (final Dosimat)

20 mL Exchange unit

10 mL Exchange unit

V-DIS = 20 mL

V-DIS = 10 mL

V-LIM = OFF

V-LIM = 15 mL

rate = 25.5 mL/min

rate = 25.5 mL/min

rate = 60 mL/min

rate = 30 mL/min


The rate