2.2. Modes

765 Dosimat

6

2.1.1 Key

The inquiries of this key are identical for all modes.

Expelling and filling rate

This key is accessible live (except in mode DOS), i.e. rate can be

changed during a running function .

ml/min

Expelling rate

Range for digital setting depending on volume of exchange unit:
1 mL 0.001

3.00 mL/min

5 mL 0.005

15.0 mL/min

10 mL

0.010

30.0 mL/min

20 mL

0.020

60.0 mL/min

50 mL

0.050

150.0 mL/min

Key sets "OFF", i.e.. the rate can be controlled

analogically by means of potentiometer at the Dosimat 765.

If the preset rate is too high to be dosed with the exchange unit

presently mounted, the rate is set automatically to its maximum.

ml/min

Filling or aspirating rate

The data input rules are the same as for the expelling rate.

The filling rate is set to maximum on changing the exchange unit (e.g.

always after

no exch. unit!

is

displayed!).

2.2 Modes

The modes are selected by the inquiry drum and

confirmed with .

DOS:

DOSing

Dosimat is dosing as long as is pressed. Result calculation

can be activated.

DIS R:

DISpensing, Repetitive

Dosimat is dosing a stored dispensing volume if is pressed,

the burette cylinder is refilled and display reset to 0.000 mL.

DIS C:

DISpensing, Cumulative

Dosimat is dosing a stored dispensing volume if is pressed,

the dispensed volume (V-DIS) remains displayed.

PIP:

PIPetting

Aspirating and subsequent expelling of a stored pipetting volume.

DIL:

DILuting

Aspirating a stored pipetting volume and subsequent expelling of the

pipetting and diluting volume.

CNT D:

CoNTent Dispenser

Preparation of solutions with preselected content.

RATE