2.2. Modes
765 Dosimat
6
2.1.1 Key
The inquiries of this key are identical for all modes.
Expelling and filling rate
This key is accessible live (except in mode DOS), i.e. rate can be
changed during a running function .
↑ ml/min
Expelling rate
Range for digital setting depending on volume of exchange unit:
1 mL 0.001
3.00 mL/min
5 mL 0.005
15.0 mL/min
10 mL
0.010
30.0 mL/min
20 mL
0.020
60.0 mL/min
50 mL
0.050
150.0 mL/min
Key
analogically by means of potentiometer at the Dosimat 765.
If the preset rate is too high to be dosed with the exchange unit
presently mounted, the rate is set automatically to its maximum.
↓ ml/min
Filling or aspirating rate
The data input rules are the same as for the expelling rate.
The filling rate is set to maximum on changing the exchange unit (e.g.
always after
no exch. unit!
is
displayed!).
2.2 Modes
The modes are selected by the inquiry drum
confirmed with
DOS:
DOSing
Dosimat is dosing as long as
can be activated.
DIS R:
DISpensing, Repetitive
Dosimat is dosing a stored dispensing volume if
the burette cylinder is refilled and display reset to 0.000 mL.
DIS C:
DISpensing, Cumulative
Dosimat is dosing a stored dispensing volume if
the dispensed volume (V-DIS) remains displayed.
PIP:
PIPetting
Aspirating and subsequent expelling of a stored pipetting volume.
DIL:
DILuting
Aspirating a stored pipetting volume and subsequent expelling of the
pipetting and diluting volume.
CNT D:
CoNTent Dispenser
Preparation of solutions with preselected content.
RATE
MODE