beautypg.com

Mode pip, pipetting – Metrohm 765 Dosimat User Manual

Page 14

background image

2.2. Modes

765 Dosimat

10

2.2.4 Mode PIP, Pipetting

Aspirating and subsequent expelling of a stored pipetting volume.

V-PIP 0.1 ml

Pipetting volume

Input range depends on the volume of the exchange unit:
1 mL

0.001... 0.900 mL

5 mL

0.001... 4.900 mL

10 mL

0.001... 9.800 mL

20 mL

0.002... 19.700 mL

50 mL

0.005... 49.500 mL

Note: The liquid of the exchange unit is mixed with the

pipetted liquid if it is aspirated into the burette cylinder!

The aspiration tube must contain V-PIP!

Expelling and filling rate see page 6.

If mode PIP is loaded, the display shows

PIP * 0.000 ml

.

The sign * in the display means that mode "PIP" is not yet ready to be used. With a first

, a preparation step is carried out which is marked in the display with

PIP prep.

. This preparation step includes the formation of an air bubble which

serves to separate the transfer solution of the exchange unit from the sample.

Then display shows

PIP 1 0.100 ml

, i.e. the Dosimat is ready to aspirate the pipetting

volume (0.1 mL). With the pipetting volume is aspirated and display shows

PIP 2 0.100 ml

which means that the Dosimat is ready to expel the pipetting volume.

With the next , the volume is expelled and the Dosimat is then ready to aspirate

the next pipetting volume without any preparation step.

If the pipetting volume is changed, a new preparation step is always carried out.

Sequence of PIP

Standard mode PIP.


Preparation step: Hold burette tip free at working

height.

Ready to aspirate the pipetting volume: Immerse

burette tip.

Ready to expel the pipetting volume: Hold

burette tip for pipetting.

Note

VOLUME

RATE

PIP * 0.000 ml

PIP prep.

PIP 1 0.100 ml

PIP 2 0.100 ml