beautypg.com

Summary, Prepare instruments for diagnostic tests, 2 summary – Metrohm 765 Dosimat User Manual

Page 26: 3 prepare instruments for diagnostic tests

background image

3.2. Diagnosis

765 Dosimat

22

3.2.2 Summary

3.2.4 Diagnosis

of cylinder code..............................................23

3.2.5 Diagnosis

of key board....................................................23

3.2.6 Diagnosis

of display ........................................................23

3.2.7 Diagnosis

analog output .................................................24

3.2.8 Diagnosis

of

digital timer .................................................25

3.2.9 Diagnosis

of

analog timer................................................26

3.2.10

Diagnosis of external inputs, outputs..............................26

3.2.11 RAM-test ..........................................................................27
3.2.12

Diagnosis of spindle zero and cock changeover............27

3.2.13 Diagnosis

of

spindle drive ...............................................27

3.2.3 Prepare instruments for diagnostic tests

1.

Power off.

2.

Remove exchange unit.

3.

Disconnect all cables at rear, except mains cable and keyboard.

4.

Power on and simultaneously press key <9> and keep pressed until

diagn. key 0...9

5.

The individual inquiries of the several diagnostic steps are accessed using the key
<0> ... <9>, exit is with the key.

diagn. key 0...9

╠══ < 0 >

╠══ < 1 >

╠══ < 2 >

╠══ < 3 >

╠══ < 4 >

╠══ < 5 >

╠══ < 6 >

╠══ < 7 >

╠══ < 8 >

╚══ < 9 >

cylinder code

keys test

display test

analog output

timer dig. test

timer ana. test

extern in/output

spind. mot. cal.

RAM TEST

RAM init.

Instrument adjustment

1)

1)

This diagnostic step is reserved for the service engineer and therefore will not be described within this document.