2.2. Modes

765 Dosimat

8

Calculation values

b = 0. ml

Blank value

Input range: 0...±999.999 mL

f = 1.

Factor

Input range: 0... ±1E33

s = 1.

Sample size

Input range: 0... ±1E33

Input manually or on-line with balance, see page 56

unit

Unit

Input range: ppm, %, g, mg, g/L, mg/L, mol, mol/L, mL, L,

/pc, none

Result calculation
If one of the calculation values (blank, factor, smpl) is not set to its standard value, a

result is calculated on filling of the Dosimat according to formula:

(dosed

volume–blank)*factor

Result = ————————————————

smpl

The result is recalculated on each entry of a calculation value (blank, factor, smpl).

Pressing display shows the dosed volume in mL. To start a new dosing, press

twice, pressing once resets the volume in display to 0.000 mL.

Printing the result on a printer
If Dosimat is set to

send RS 232 on

(special settings see page 18), filling the Dosimat or

re-calculation triggers a print command. A continuous number (#), the dosed volume and

the calculated result are printed.

Do set new calculation parameters for the next dosing only after having pressed

once, i.e. if display shows

DOS 0.000 ml

.

The continuous number is set to zero on switching on the Dosimat and incremented by 1

on every filling command.

Example of print-out

7

BLANK

SMPL

8

FACTOR

UNIT

9

4

#01 V = 0.352 ml R = 7.04 ppm
#02 V = 0.440 ml R = 8.8 ppm

#03 V = 0.000 ml

#04 V = 0.364 ml R = 7.28 ppm

#05 V = 0.438 ml R = 8.76 ppm

#06 V = 0.382 ml R = 7.64 ppm

#07 V = 0.370 ml R = 19.61 %

#08 V = 0.372 ml R = 19.72 %