4.2 Control commands

765 Dosimat

41

The entry is automatically corrected to a multiple of 1/10 000 V(B) depending on the

volume of the cylinder:

Cylinder

ml

Volume

min

ml/min

Volume

max

ml/min

1 0.001

0.900

5 0.001

4.9009

10 0.001

9.800

20 0.002

19.700

50 0.005

49.500

If a value is entered which is too high or too low, resp. , the Dosimat corrects it

automatically to volume

max

or volume

min,

resp. and sets bit 1 of information byte 2 to 1 (see

4.2.7).

4.2.30 VDL VALUE

Diluting volume, not live

Value = 0.001 ... 999.999

Setting the diluting volume in ml.

In mode DIL only, in other modes the command is not accepted and bit 0 of information

byte 2 is set to 1.

The entry is automatically corrected to a multiple of 1/10 000 V(B) depending on the

volume of the cylinder:

Cylinder

ml

Volume

min

ml/min

Volume

max

ml/min

1 0.001

999.999

5 0.001

999.999

10 0.001

999.999

20 0.002

999.999

50 0.005

999.999

If a value is entered which is too high or too low, resp. , the Dosimat corrects it

automatically to volume

max

or volume

min,

resp. and sets bit 1 of information byte 2 to 1 (see

4.2.7).

4.2.31 VLI VALUE

Security volume, not live

Value = 0.001 ... 999.999

Setting the security volume in ml.

In modes DOS, DIS C, and PULSE only, in other modes the command is not accepted

and bit 0 of information byte 2 is set to 1.

The entry is automatically corrected to a multiple of 1/10 000 V(B) depending on the

volume of the cylinder: