4.2 Control commands
765 Dosimat
41
The entry is automatically corrected to a multiple of 1/10 000 V(B) depending on the
volume of the cylinder:
Cylinder
ml
Volume
min
ml/min
Volume
max
ml/min
1 0.001
0.900
5 0.001
4.9009
10 0.001
9.800
20 0.002
19.700
50 0.005
49.500
If a value is entered which is too high or too low, resp. , the Dosimat corrects it
automatically to volume
max
or volume
min,
resp. and sets bit 1 of information byte 2 to 1 (see
4.2.7).
4.2.30 VDL VALUE
Diluting volume, not live
Value = 0.001 ... 999.999
Setting the diluting volume in ml.
In mode DIL only, in other modes the command is not accepted and bit 0 of information
byte 2 is set to 1.
The entry is automatically corrected to a multiple of 1/10 000 V(B) depending on the
volume of the cylinder:
Cylinder
ml
Volume
min
ml/min
Volume
max
ml/min
1 0.001
999.999
5 0.001
999.999
10 0.001
999.999
20 0.002
999.999
50 0.005
999.999
If a value is entered which is too high or too low, resp. , the Dosimat corrects it
automatically to volume
max
or volume
min,
resp. and sets bit 1 of information byte 2 to 1 (see
4.2.7).
4.2.31 VLI VALUE
Security volume, not live
Value = 0.001 ... 999.999
Setting the security volume in ml.
In modes DOS, DIS C, and PULSE only, in other modes the command is not accepted
and bit 0 of information byte 2 is set to 1.
The entry is automatically corrected to a multiple of 1/10 000 V(B) depending on the
volume of the cylinder: