beautypg.com

Continuous dosing with two dosimats, 5 continuous dosing with two dosimats – Metrohm 765 Dosimat User Manual

Page 61

background image

5.5 Continuous dosing with two Dosimats

765 Dosimat

57

5.5 Continuous dosing with two Dosimats

Two 765 Dosimats are suitable for continuous dosing.

The two 765 Dosimats are connected with cable 3.980.3140 via both interfaces RS232 (6).
Procedure
• Select mode DIS C in both Dosimats.
• Set on both Dosimats the dispensing volume equal to the volume of the mounted

exchange unit (V-DIS =

V

burette

).

If one of the Dosimats is started and reaches the preselected dispensing volume, this

triggers the start of the other Dosimat, etc.

• Prerequisite for continuous dosing is that the filling rate is higher than the expelling rate

(while one Dosimat is expelling, the other one is refilling i.e. in addition to the

movement of the piston two turns of the cock are necessary, about 2 s for each

turning).
Example: rate ↑ ≤ 0.85 rate ↓

max

for the smaller of the exchange unit

At high rates, the effective expelling rate can deviate from the set rate by max. 4 % as a

result of the finite resolution of the digital rate control, but the effective rate is held

constant with quartz crystal accuracy.
If dosing to a fixed end volume V is envisaged, V-LIM is set according to the following

rules:
• Calculation of the number of complete cycles N (V

1

and V

2

be completely expelled):

V

N = INT ———— (integral quotient)

V

1

+ V

2

• Calculation of the remaining volume V

R

:

V

R

= V - N

*

(V

1

+ V

2

)

• In case the remaining volume is V

R

=0, Dosimat 2 is the final burette.

Settings on Dosimat 1:

V-LIM(1) =

OFF

VDIS(1) =

V

1

Settings on Dosimat 2:

V-LIM(2)

=

N

*

V

2

V-DIS(2)

=

V

2

• In case the remaining volume is V

R

≠0 and V

R

≤ V

1

, Dosimat 1 is the final burette.

Settings on Dosimat 1:

V-LIM(1) = V - N * V

2

VDIS(1) =

V

1

Settings on Dosimat 2:

V-LIM(2)

=

OFF

V-DIS(2)

=

V

2

• In case the remaining volume is V

R

≠0 and V

R

> V

1

, Dosimat 2 is the final burette.

Settings on Dosimat 1:

V-LIM(1) =

OFF

VDIS(1) =

V

1

Settings on Dosimat 2:

V-LIM(2) = V - V

1

* (N + 1)

V-DIS(2)

=

V

2

with

V

1

cylinder volume on Dosimat 1 (start Dosimat)

V

2

cylinder volume on Dosimat 2