Metrohm 765 Dosimat User Manual
Page 42
4.2. Control commands
765 Dosimat
38
4.2.15 MDC
Mode DIS C with previous parameters, not live
Selection of mode DIS C without changing the actual parameters in the working memory.
No filling of the exchange unit.
4.2.16 MPU ON
Mode PULSE on, not live
Mode PULSE on.
Mode PULSE is not equivalent to modes DOS, DIS R, DIS C, PIP or DIL. Mode PULSE
runs before one of these modes. The parameters in the working memory remain
unchanged and no filling is executed.
In mode PULSE, 1/10 000 V(B) is dosed with every 'GO'. The maximal frequency of GO-
commands is ≤ 500 Hz. This is equal to the maximum rate of 1 V(B)/20 s.
Note: It is advantageous to run mode PULSE before mode DOS or DIS C, because in
these two modes V-LIM is active. Mode PULSE has to be left with "MPU OFF".
4.2.17 MPU OFF
Mode PULSE off, not live
Mode PULSE off. The background mode is reactivated in the working memory. No
parameters are changed and no filling is executed.
4.2.18 MST X
Mode store under address X, not live
X = 0... 9, J
Storing the actual mode with the corresponding parameters under address X. The content
of the working memory remains unchanged.
Store J is only accessible via remote control and serves e.g. to store the actual mode
after "REM ON".
4.2.19 MRC X
Mode recall with address X, not live
X = 0 ... 9, J
Recalling a mode with the corresponding parameters from the user memory.
Mode CNT D is not loaded into the working memory and bit 0 of information byte 2 is set
to 1.
4.2.20 PBL VALUE
Parameter 'blank', live
Value = -999.999 ... +999.999
Setting blank in ml.
In mode DOS only, in other modes, the command is not accepted and bit 0 of information
byte 2 is set.
4.2.21 PFA VALUE
Parameter 'factor', live
Value = -1E33 ... -1E-37 , 0 , 1E-37 ... 1E33
Setting factor.
In mode DOS only, in other modes, the command is not accepted and bit 0 of information
byte 2 is set to 1.