Page 27: 4 diagnosis, Of cylinder code, 5 diagnosis, Of key board, 6 diagnosis, Of display, 4 diagnosis of cylinder code

background image

3.2 Diagnosis

765 Dosimat

23

3.2.4 Diagnosis of cylinder code

1.

Prepare instrument for diagnostic test (see chap. 3.2.3).

2.

<0>

cylinder code

3.

no exch. unit!

4.

Insert (dummy) exchange unit.

code: xx ml

5.

Check whether the displayed ml-Code (xx ml) corresponds to the exchange unit.

Various exchange units can be inserted to verify their ml-code.

If an exchange unit is coded incorrectly or if the code switches are inoperative, the display shows: E 90: ..

no code!

6.

diagn. Key 0...9

3.2.5 Diagnosis of key board

1.

Prepare instrument for diagnostic test (see chap. 3.2.3).

2.

<1>

keys test }

3.

key: rate

The display requests to press the key on the keyboard 6.2149.000.

4.

, <7>, <4>, <1> etc.

If the correct key was pressed and the test was positive the name of the next key to be pressed will be dis-

played.

Display E 10: and on the right-hand side the name of an other key indicates a fault in the keyboard matrix, or

the wrong key was pressed.

A fault indication may be cancelled by pressing . “breaking off ?“ is then displayed, asking you

whether you want to stop the test or not.

- To stop press .

- To continue press until the display shows “keys test end“

After pressing the last key (GO), the following appears:

keys o.k.

5.

diagn. key 0...9

3.2.6 Diagnosis of display

1.

Prepare instrument for diagnostic test (see chap. 3.2.3).