Metrohm 765 Dosimat User Manual
Page 41
4.2 Control commands
765 Dosimat
37
Information byte 2:
Bit Function
0
1= Wrong command code
1
1 = Parameter corrected to its limit value
2
1 = Repeat command in the READY state
3
1 = Cylinder empty
4
1 = Remote control on
5
1 = Data transfer on (send RS232 on)
6 Reserve
4.2.8 DOS
Mode DOS, not live
Selection of standard mode DOS. The standard parameters (depending on the exchange
unit) are loaded into the working memory. If the exchange unit is not filled, filling is
executed.
4.2.9 DIS R
Mode DIS R, not live
Selection of standard mode DIS R. The standard parameters (depending on the
exchange unit) are loaded into the working memory. If the exchange unit is not filled,
filling is executed.
4.2.10 DIS C
Mode DIS C, not live
Selection of standard mode DIS C. The standard parameters (depending on the
exchange unit) are loaded into the working memory. If the exchange unit is not filled,
filling is executed.
4.2.11 PIP
Mode PIP, not live
Selection of standard mode PIP. The standard parameters (depending on the exchange
unit) are loaded into the working memory. If the exchange unit is not filled, filling is
executed.
4.2.12 DIL
Mode DIL, not live
Selection of standard mode DIL. The standard parameters (depending on the exchange
unit) are loaded into the working memory. If the exchange unit is not filled, filling is
executed.
4.2.13 MDO
Mode DOS with previous parameters, not live
Selection of mode DOS without changing the actual parameters in the working memory.
No filling of the exchange unit.
4.2.14 MDR
Mode DIS R with previous parameters, not live
Selection of mode DIS R without changing the actual parameters in the working memory.
No filling of the exchange unit.