Mode cnt d, content dispenser – Metrohm 765 Dosimat User Manual

Page 16

background image

2.2. Modes

765 Dosimat

12

Sequence of DIL

Standard mode DIL.

Preparation step: Hold burette tip free at working

height.

Ready to aspirate the pipetting volume: Immerse

burette tip.

Ready to expel the pipetting and diluting volume:

Hold burette tip for pipetting.

Notes
• If you wish to change V-PIP, it is best to change it during filling in the preparation step,

i.e. when display shows

DIL prep.

.

• If V-PIP is changed at another time, a new preparation step is carried out, which

changes the volume of the air bubble. The first dilution after such a change could be

erroneous and should be discarded. Another possibility is to expel the air bubble in

mode DOS and to start Mode DIL from the beginning. V-DIL can be changed at any

time without a new preparation step.

2.2.6 Mode CNT D, Content Dispenser

Mode CNT D is used to prepare solutions with a particular content. Doing this, the sub-

stance must not be weighed-out to a particular value in order to obtain the preselected

content but the 765 Dosimat dispenses the amount of solvent calculated correspond-

ingly.

Expelling and filling rate see page 6.

RATE

DIL * 0.000 ml

DIL prep.

DIL 1 0.100 ml

DIL 2 1.100 ml

DIL prep.