Mode dos, dosing – Metrohm 765 Dosimat User Manual

Page 11

background image

2.2 Modes

765 Dosimat

7

Example:

Selection of mode "DIS C", cumulative dispensing.

Press .

Display shows that mode which has been selected last with key , e.g.

DOS .

Press repeatedly until display shows

DIS C

.

Load mode "DIS C" into working memory with .

Display shows

DIS C 0.000 ml

.

Mode "DIS C" is ready to work, the piston is in zero position.

All modes which are loaded into the working memory by key are equipped

with a set of standard parameters:

Mode V-DIS/V-PIP

mL

V-LIM/V-DIL

mL

Rate

mL/min

Rate

mL/min

Calculation

DOS

OFF

OFF

max.

b=0; f=1; s=1

DIS R

1

OFF

max.

DIS C

0.1

OFF

OFF

max.

PIP 0.1 – OFF

OFF

DIL 0.1 1 OFF

OFF

CNT D

OFF

max.

2.2.1 Mode DOS, Dosing

Dosimat is dosing as long as is pressed. Result calculation can be activated.

V-LIM OFF ml

Security volume:

Dosing is stopped if V-LIM has been reached.

Input range: 0.001...999.999 mL, OFF

Expelling and filling rate, see page 6.

VOLUME

RATE