2.2 Modes

765 Dosimat

11

• A new air bubble is built with every preparation step ("prep.") e.g. its volume increases.

If you wish to keep the volume of the air bubble expel it in mode DOS before changing

V-PIP.

• For best pipetting results we recommend exchange units with volumes ≤ 20 mL and

pipetting equipment 6.5611.000.

• The aspirating and expelling rates should not be higher than 20 mL/min.

• Hold tubing tip in an angle of app. 45° to the vessel wall during pipetting. Just the

same as you do with glass pipettes!

• The vessel, containing the liquid you want to pipette should stand on the same level as

the vessel into which you are going to expel the liquid in order to ascertain app. the

same level of the pipetting tubing during work.

2.2.5 Mode DIL, Diluting

Aspirating a stored pipetting volume and subsequent expelling of the pipetting and dilut-

ing volume.

V-PIP 0.1 ml








V-DIL 1. ml

Pipetting volume

Input range depends on the volume of the exchange unit:
1 mL

0.001... 0.900 mL

5 mL

0.001... 4.900 mL

10 mL

0.001... 9.800 mL

20 mL

0.002... 19.700 mL

50 mL

0.005... 49.500 mL

Diluting volume

Input range: 0.001...999.999 mL

Note: The diluting liquid is unintentionally mixed with the

pipetted liquid if it is aspirated into the burette cylinder!

The aspiration tube must contain V-PIP!

Expelling and filling rate see page 6.


If mode DIL is loaded, the display shows

DIL * 0.000 ml

.

The sign * in the display tells you that mode DIL is not ready to be used. With a

preparation step is carried out during which V-PIP is expelled into the bottle of the

exchange unit and an air bubble is built to separate the solution of the exchange unit from

the sample.
Then the Dosimat is ready to aspirate the pipetting volume (0.1 mL) which is displayed by

DIL 1 0.100 ml

and carried out after pressing .

Then

DIL 2 1.100 ml

is displayed which means that the Dosimat is ready to expel the

pipetting and the diluting volume (0.1 mL + 1 mL = 1.1 mL). This is executed after

. The preparation step is now carried out automatically and the Dosimat is ready to

aspirate the next pipetting volume.

VOLUME

RATE