4.2. Control commands
765 Dosimat
44
4.2.46 QAD
Query rate down 'analogue on/off', live
Request to send the information rate down 'analogue on/off' via RS232 interface.
For example:
Analogue on "on 'CR' 'LF'"
Analogue off " off 'CR' 'LF'"
4.2.47 QDS
Query dispensing volume, live
Request to send the dispensing volume via RS232 interface.
For example: "1.275 'CR' 'LF'"
Transmission is only possible in modes DIS R and DIS C.
In other modes, string "not defined 'CR' 'LF'" is transmitted.
4.2.48 QPI
Query pipetting volume, live
Request to send the pipetting volume via RS232 interface.
For example: "1.275 'CR' 'LF'"
Transmission is only possible in modes PIP and DIL.
In other modes, string "not defined 'CR' 'LF'" is transmitted.
4.2.49 QLI
Query limit volume, live
Request to send the limit volume via RS232 interface.
For example: "1.275 'CR' 'LF'"
If V-LIM is 'off', the string "OFF 'CR' 'LF'" is transmitted.
Transmission is only possible in modes DOS and DIS C.
In other modes, string "not defined 'CR' 'LF'" is transmitted.
4.2.50 QDL
Query diluting volume, live
Request to send the diluting volume via RS232 interface.
For example: "1.275 'CR' 'LF'"
Transmission is only possible in mode DIL.
In other modes, string "not defined 'CR' 'LF'" is transmitted.
4.2.51 QUN
Query unit, live
Request to send the unit via RS232 interface.
For example: "mg/l 'CR' 'LF'"
Transmission is only possible in mode DOS.
In other modes, string "not defined 'CR' 'LF'" is transmitted.
4.2.52 QAF
Query 'auto fill on/off', live
Request to send the information 'auto fill on/off' via RS232 interface.
For example:
auto fill on "on 'CR' 'LF'"
auto fill off "off 'CR' 'LF'"