5.1 Technical specifications
765 Dosimat
51
5 Appendix
5.1 Technical specifications
Exchange units
1, 5, 10, 20, 50 mL burette cylinder volumes,
preferably with flat cock for automatic cock changeover
Resolution
10'000 pulses per 100% of burette volume
Exactitude
Metrohm dosimats and exchange units meet the requirements
of ISO/EN/DIN Standard 8655-3 "Piston-operated volumetric
apparatus – Part 3: Piston burets" and DIN Standard 12 650.
Limits according to ISO/EN/DIN 8655-3
Cylinder
volume
Max. permissible system. error
Max. permissible random error
1 mL
± 0.6%
± 6 µL
± 0.1%
± 1 µL
5 mL
± 0.3%
± 15 µL
± 0.1%
± 5 µL
10 mL
± 0.3%
± 20 µL
± 0.07%
± 7 µL
20 mL
± 0.2%
± 40 µL
± 0.07%
± 14 µL
50 mL
± 0.2%
± 100 µL
± 0.05%
± 25 µL
Metrohm agencies throughout the world offer you the possibil-
ity of checking the accuracy of your exchange units and Dosi-
mats locally and also of certifying them. If the dosing cylinder
and/or piston of an exchange unit are replaced then we rec-
ommend that an accuracy check is carried out.
Dispensing time for 100% of burette cylinder volume
analogue setting
20 s ... app. 17 min
digital setting
20 s ... app. 17 h
Modes
DOS
Dosing, with the ability to calculate a result from the dosed
volume
DIS R
Repetitive Dispensing
DIS C
Cumulative Dispensing
PIP
Pipetting
DIL
Diluting
CNT D
Preparation of solutions with preselected content
User memory
for 10 complete user modes
Data outputs
digital
Data transfer interface according to RS 232 C
analogue
Representation
Volume as voltage signal: 0 ... 1000 mV = 1, 2 ... 10 burette
cylinder volumes (selectable)
Resolution
1
mV