5.1 Technical specifications

765 Dosimat

51

5 Appendix

5.1 Technical specifications

Exchange units

1, 5, 10, 20, 50 mL burette cylinder volumes,

preferably with flat cock for automatic cock changeover

Resolution

10'000 pulses per 100% of burette volume

Exactitude

Metrohm dosimats and exchange units meet the requirements

of ISO/EN/DIN Standard 8655-3 "Piston-operated volumetric

apparatus – Part 3: Piston burets" and DIN Standard 12 650.

Limits according to ISO/EN/DIN 8655-3

Cylinder

volume

Max. permissible system. error

Max. permissible random error

1 mL

± 0.6%

± 6 µL

± 0.1%

± 1 µL

5 mL

± 0.3%

± 15 µL

± 0.1%

± 5 µL

10 mL

± 0.3%

± 20 µL

± 0.07%

± 7 µL

20 mL

± 0.2%

± 40 µL

± 0.07%

± 14 µL

50 mL

± 0.2%

± 100 µL

± 0.05%

± 25 µL

Metrohm agencies throughout the world offer you the possibil-

ity of checking the accuracy of your exchange units and Dosi-

mats locally and also of certifying them. If the dosing cylinder

and/or piston of an exchange unit are replaced then we rec-

ommend that an accuracy check is carried out.

Dispensing time for 100% of burette cylinder volume

analogue setting

20 s ... app. 17 min

digital setting

20 s ... app. 17 h

Modes

DOS

Dosing, with the ability to calculate a result from the dosed

volume

DIS R

Repetitive Dispensing

DIS C

Cumulative Dispensing

PIP

Pipetting

DIL

Diluting

CNT D

Preparation of solutions with preselected content

User memory

for 10 complete user modes

Data outputs

digital

Data transfer interface according to RS 232 C

analogue

Representation

Volume as voltage signal: 0 ... 1000 mV = 1, 2 ... 10 burette

cylinder volumes (selectable)

Resolution

1

mV