Pc-2200, 2 application examples 6.3 diameter/thickness a b – Bowers Group Advanced Electronic/Air Column PC-2200 User Manual
Page 29

PC-2200
BOWERS METROLOGY page 29 /32 version 2.06 -E
6.2 APPLICATION EXAMPLES
6.3 DIAMETER/THICKNESS
A
B
Menu
Sub-menu
Selection
Function
Choice
Values
CHANNEL
SOURCE EXT.
EXT.
CH-BUS
CH-A
RATIO
0 - 5.000
1.000
DIREC.
+ / -
+
CH-B
RATIO
0 - 5.000
1.000
DIREC.
+ / - +
CH-C
RATIO
0 - 5.000
1.000
DIREC.
+ / -
+
CH-D
RATIO
0 - 5.000
1.000
DIREC.
+ / -
+
MEASURE
M1 SELECT
ON
ON
BAR
BOTTOM
CENTER
-10->0
CENTER
B.FACT
-5.00 / +5.00
1.00
RANGE
± 1.0mm
± 0.5mm
± 0.1mm
± 0.05mm
± 0.01mm
± 0.005mm
± 0.05mm
SENS.
± 0.01 ~ 5.000
1.000
RESOL.
0.01, 0.005, 0.001,
0.0005, 0.0002,
0.0001
0.001
FUNC
A, B, C, D, A+B, A-B,
C+D, (A+B)-C,
(A+B+C)/3,
(A+B+C+D)/4
(A+B)-(C+D)
A+B
MODE
STATIC,
MAXIMUM,
MINIMUM,(M+m)/2,
(M-m)/2, M-m
STATIC
SPEC.
M1 NOMI.
000.0000
010.000
-TOLER.
-0.8000
-0.0200
+TOLER
+0.0000
+0.0200
-APPRO
-0.0000
-0.0150
+APPRO
+0.0000
+0.0150
MASTER
M1 -MASTER
+00.0000
+9.9800
+MASTER
+00.0000
+10.0200
CLASS
SELECT
OFF / ON
OFF
NUMBER
1 - 100
010
START
+00.0000
+00.0000
STEP
0.0000
0.0100
RS232
SPEED
19200, 9600, 4800
9600
DATA
7,
8
8
PARITY
None,even,Odd
None
TERM
CR,CRLF
CRLF
RTS
ON / OFF
ON
BRIEF,
FULL
BRIEF
RSMODE
PRINT,HOST
ID
NO.
000
000
ETC
GENERAL UNIT
METRIC,INCH
METRIC
DISPLAY
mm,
um
mm
INPUT
HBT,LVDT,CAP,AIR
HBT
P/W
PROGRAM
0000
CALIBR.
ON/OFF
OFF
AVERAGE
BAR
000
003
DIGIT
000
003
A/D
00
10
VERSION
V2.00
External diameter : 10.000 mm
Lower tolerance : - 20
µm
Upper tolerance: +20
µm
Lower warning limit: - 15
µm
Upper warning limit: +15
µm
Lower master : 9.980 mm
Upper master : 10.020 mm
Unit : mm