beautypg.com

Pc-2200 – Bowers Group Advanced Electronic/Air Column PC-2200 User Manual

Page 14

background image

PC-2200

BOWERS METROLOGY page 14 /32 version 2.06 -E



3.4 PROGRAMMING A DIAMETER

Measurement of a diameter with a HBT probe connected to channel A.




























Step n°1 is to place the probe into a location to be able to get the maximum range and the minimum
deviation.
Place the measuring head of 15.0000 mm in the master,
Press simultaneously

Move the probe down to the electro-mechanical zero. , then approximately stop around

zero, lock the probe . Press to escape to the measuring mode.







Parameters to be changed for example number 2

CHANNEL : Direc. -
MEASURE : M1 –Range :

± 0.005 mm Resol.: 0.0001mm

SPEC. : M1 – NOMI. 015.0000 - -TOLER –0.0020 - +TOLER +0.0020 –APPROCH –0.0016
+ APPROCH : +0.0016
MASTER : M1 - -Master : 14.9988mm - +Master : 15.0026mm

Component

15.000 mm ± 0.002

Min master
14.9988mm

Max. master.
15.0026mm

S

Enter

Enter

Esc

A D J U S T

-

0 . 0 0 1 0

C H - A

- 1 . 0 5 0 0