3.2 PROGRAMMING A SIMPLE MEASURE
Example N°1
Measure a component with a HBT probe connected to channel A.
Step n°1 is to place the probe into a correct position to be able to get the maximum range and the minimum
deviation.
Place the master 20.0000 mm on the stand, place the probe into position.
Press simultaneously
Move the probe down to the electro-mechanical zero, , then approximately stop around
zero, lock the probe . Press to escape to the measuring mode.
Parameters to be changed for example number 1
MEASURE : M1 –Range :
± 0.01mm Resol.:0.0002mm –
SPEC.
: M1 – NOMI. 020.0000 - -TOLER –0.0050 - +TOLER +0.0050 –APPROCH –0.0040
+
APPROCH
:
+0.0040
Master
20.0000mm
Component
20.000mm ± 0.005
USING a HBT probe, we do not use the
calibration, this mode will be explained in a
subsequent chapter.