Pc-2200 – Bowers Group Advanced Electronic/Air Column PC-2200 User Manual
Page 11
PC-2200
BOWERS METROLOGY page 11 /32 version 2.06 -E
Chapter 3. HOW TO USE IT
3.3 SWITCHING
ON THE COLUMN
To start the first time, we advise to make a general RESET .
Press + ON simultaneously,
Do not release your pressure on before the message RESET appears on the display, then a test
program will start.
The default parameters are :
CHANNEL : Source : Ext. - CH-A Ratio :1.000 Direc. +
MEASURE : M1 – Bar :Bottom B.fact. 1.00 – Range :
± 0.05” Sens : 1.000 Resol.:0.00005” – Func: A –
Mode: STATIC
SPEC. : M1 – NOMI. 000.0000 - -TOLER –0.0500 - +TOLER +0.0500 –APPROCH –0.0500
+ APPROCH : +0.0500
MASTER : M1 - -MASTER +0.00000 +MASTER +0.00000
CLASS : OFF
RS232 : SPEED 9600, DATA 8, PARITY N, TERM CRLF, RTS ON, PRINT BRIEF, RSMODE PRINT
ID NO. 000
ETC : GENERAL UNIT INCH, DISPLAY mm, INPUT HBT, P/W PROGRAM 0000, CALIBR OFF
AVERAGE BAR 005, DIGIT 005, A/D 20
VERSION V2.xx
The above parameters are the default values, so it is necessary to modify the values according to your
application.
Description of functions per menu
CHANNEL
: Select the source and the probes used with ratio and measuring direction.
MEASURE
: Select the starting position of the bar, Bar factor, Range of the Bar, Resolution of the digital
display, Combination of probes, Static or Dynamic measuring mode.
SPEC.
: Store the nominal size, the upper tolerance, the lower tolerance, the upper warning limit, the
lower warning limit of the component to be measured.
MASTER
: Store the lower master value and the upper master value used.
CLASS
: Active the classification mode, class number, starting value, class range.
RS232
: Select the transmission speed, word length, parity control, end of characters, request to send
signal, printing mode, simplex or duplex mode, identity number.
ETC
GENERAL : Select the unit (mm or inch), type of display (mm or um), type of probes used, program and
calibration password.
AVERAGE : Select the bar sensitivity , the display sensitivity, the analog board sensitivity.
VERSION : Show the software version.
Enter
▓ B O W E R S ▓
R E S E T . . .
- 1 . 0 4 0 0
Enter