RKI Instruments Eagle 2 User Manual
Page 83
Changing Eagle 2 Instrument Parameters • 79
the standard four gas section.
6. To change the serial number, click the serial number field
and use the backspace key to remove the current entry, then
type the new information. To change the station ID or user
ID, click the drop-down menu and select the desired station
or user ID.
7. To change the datalogging trend interval time, click on the
down arrow in the Interval Trend Time Field and select the
desired interval time in seconds. The available choices are
5, 10, 20, 30, 60, 180, 300, and 600 seconds.
8. To change the channel parameters, click the field you wish to
change (e.g. H2S AutoCal.) to select it, then type the new
information.
9. After you have finished entering new parameters, you must
upload this information to the Eagle 2 by clicking the Update
button, then confirming that you want to update the
information by clicking the Yes button when the Update
window appears.
10.To update the Eagle 2’s date and time to match the
computer’s, click the Date/Time Set button and then click
the Yes button when the Update window appears.
11.If you wish to use other functions in the program, click the
appropriate control button along the right side of the program
window.
12.If you wish to exit the software, click the Download control
button to return to the Download Window. Turn off the Eagle
Figure 65: Update Window