66 • Viewing, Printing, and Deleting Data in the Last Calibration Window
Viewing, Printing, and
Deleting Data in the Last
Calibration Window
You can access data on the most recent successful calibration
for each instrument that has been downloaded in the Last
Calibration Window. You can view, print (calibration date view
option only), and delete this data in the Last Calibration Window.
Viewing and Printing Last Calibration Data
Open the Last Calibration Window by clicking Last Calibration
along the right side of the program window. When you open the
Last Calibration Window the first time after launching the
program, it will open with the Need Calibration view option
selected. There are three view options in the Last Calibration
window: Need Calibration, Calibration Date, and Calibration
Record.
Need Calibration Option
Selecting this option shows the last calibration date and the last
download date for the Eagle 2s that are due for calibration.
NOTE: The calibration interval, the number of days after a
calibration that a new calibration is due, is saved in the
Eagle 2. If the calibration interval is changed in the
Eagle 2, the new calibration interval will not be known
by the program until the Eagle 2 is downloaded.