96 • Detail Settings Button
from Table 5 on page 102 or through testing as described in
“Obtaining a Relative Response Factor” on page 101. A valid
character for the Factor is a value between 0.01 and 25.00.
There are limitations for the full scale and increment values that
depend on the factor for both the high range and the low range.
Table 2 and Table 3 below list these limitations.
Table 2: High Range PID
Factor
Full Scale
(ppm)
Increment
0.25-0.36
1000
1
0.37-0.49
1500
1
0.50-0.61
2000
1
0.62-0.74
2500
1
0.75-0.99
3000
2
1.00-1.24
4000
2
1.25-2.49
5000
2
2.50-3.74
10000
10
3.75-4.99
15000
10
5.00-7.49
20000
10
7.50-9.99
30000
20
10.00-
12.49
40000
20
12.50
50000
2500
Table 3: Low Range PID
Factor
Full Scale
(ppm)
Increment
0.20-0.29
10.00
0.01
0.30-0.39
15.00
0.01
0.40-0.49
20.00
0.01