WARNING! Always pull out the supply plug before

opening the oven cover!

4.3 Electrical equipment

Operating voltage see label 50/60 Hz

Current rating see label

Protection Class 1, i.e. operating isolation with ground connection to EN 61 010

Protection IP20 to DIN EN 60 529

Interference suppression to EN55011 Class B

Oven protected by a fuse 250V/15A fast blow

Controller protected by a 100 mA fuse (200 mA on 115 V)

When connecting a MEMMERT oven to the electrical supply you have to observe any local

regulations which apply (e.g. in Germany DIN VDE 0100 with FI protection circuit)

This product is intended to operate on a supply network with a system impedance Zmax at the

transfer point (building connection) of 0.292 Ohm max. The user has to ensure that the product is

only operated on an electrical supply network which meets these requirements. If necessary, details

of the system impedance can be obtained from the local electricity supply authority.

Note:

Any work involving opening up the oven must only be carried out by a suitably qualified

electrician!

4.4 External

connection

Equipment connected to the external connections must have interfaces which meet the requirements for
safe low voltage (e.g. PC, printer).