19 Sterilisation chip card (for INP incubators only)

Sterilisation

Process Control

Card

Sterilisation Parameter:

160 °C / 4:00 h

Unit-ID:

________________________

The STERICard starts an automatic and preset sterilisation process.

The automatic sterilisation process starts as soon as the STERICard is inserted into the oven and started by
the user.

Ramp 1:

heating to 160°C

Ramp 2:

hold for 4 hours (after the sterilisation temperature has been reached)

Ramp 3:

cooling to 70°C

Ramp 4:

wait 10 minutes

After successful sterilisation the control displays STERILISATION OK.
The air valve remains closed during the entire sterilisation sequence.
The automatic sterilisation programme can not be altered by the user. For documentation purposes the
sterilisation process can be reported on the STERICard and can be read using „Celsius“.

Important note:

On incubators the sterilisation programme does not serve for

sterilising the load but solely for sterilising the chamber. The incubator

is not a steriliser within the scope of the medical product regulations.