Muster – Memmert IFP 400 - 800 Incubators (Generation 2003) User Manual
Page 7

PERFECT
Muster
page 7
4 Technical
data
supply plug
connection
RS232C
printer
connection
Model
200
300
400
500
550
600
700
800
Chamber width A [mm]
400
480
400
560
480
800
1040
1040
Chamber height B [mm]
320
320
400
480
640
640
800
1200
Chamber depth C [mm]
250
250
330
400
500
500
500
600
Oven width D [mm]
550
630
550
710
630
950
1190
1190
Oven height E [mm]
600
600
680
760
920
920
1080
1605
Oven depth F [mm]
400
400
480
550
650
650
650
750
Chamber volume [litre]
32
39
53
108
153
256
416
749
Weight [kg]
28
30
35
50
82
87
121
170
Power, ovens UNP/UFP/SFP [W]
1100
1200
1400
2000
2200
2400
4000
4800
Power, ovens INP/IFP [W]
440
500
800
900
1100
1600
1800
2000
Max. number of shelves
3
3
4
5
7
7
9
14
Max. load per shelf [kg]
30
30
30
30
30
30
30
30
Max. load per oven [kg]
30
30
90
60
80
80
100
160
Ambient conditions
Ambient temperature 5°C to 40°C
rH 80% max., no condensation
Overvoltage category: II
Contamination level: 2
Setpoint temperature range
20°C to nominal temperature (details see label)
Setting accuracy:
up to 100°C: 0.1°C
from 100°C: 0.5°C
Working temperature range
From 5°C above ambient temperature up to nominal temperature =
maximum temperature (details see label)
With fan switched on (UFP/SFP) from 10°C above ambient temperature
up to nominal temperature = maximum temperature (details see label)