Memmert IFP 400 - 800 Incubators (Generation 2003) User Manual
Page 46

PERFECT
page 46
14. Select duration of third ramp segment
Using the push/turn control select the time indication
Hold down the SET key and set the time 00:01 using the push/turn
control
h
15. Select temperature of third ramp segment
Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of
30 °C using the push/turn control.
˚C
16. Select fan speed for third ramp segment
Turn the push/turn control clockwise until the fan symbol is flashing.
Hold down the SET key and set the fan speed to 100% using the push/
turn control.
17. Set air valve for third ramp segment
Turn the push/turn control clockwise until the air valve symbol is flashing.
Hold down the SET key and set the air valve to 100% using the push/
turn control.
18. Set closure command for third ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g. end ) appears.
Hold down the SET key and set SPWT [T] with the push/turn control.
The segment closure command
SPW [T] ensures that
sterilisation is terminated
only after the unloading
temperature has been reached.
19. Select duration of fourth ramp segment
Using the push/turn control select the time indication
Hold down the SET key and set the time 00:01 using the push/turn
control
h
20. Select temperature of fourth ramp segment
Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of
20 °C using the push/turn control.
˚C
21. Select fan speed for fourth ramp segment
Turn the push/turn control clockwise until the fan symbol is flashing.
Hold down the SET key and set the fan speed to 100% using the push/
turn control.