Muster – Memmert IFP 400 - 800 Incubators (Generation 2003) User Manual
Page 27

PERFECT
Muster
page 27
13.2 Electronic temperature monitor
13.2.1 Overtemperature protection
MAX
Range: up to 10°C max above nominal
temperature (for nominal temperature
see label)
MIN
MAX
AUTO
˚C
Using the push/turn control select the
symbol
MAX
-Symbol anwählen.
Hold down the SET key and set the
protection temperature using the
push/turn control.
13.2.2 Undertemperature protection
MIN
Range: from 10°C below minimum
temperature of oven to 10°C above
nominal temperature of oven (for
nominal temperature see label).
The low alarm cannot be programmed
above the value set as high alarm.
Where no undertemperature
protection is required, this has to be
set to the lowest temperature.
MIN
MAX
AUTO
°C
Using the push/turn control select the
symbol
MIN
.
Hold down the SET key and set the
protection temperature using the
push/turn control.
Note:
The temperature monitor can be set independently of the operating mode.
During ramp operation the monitor temperature must always be set sufficiently far above the maximum
working temperature.