13.2 Electronic temperature monitor

13.2.1 Overtemperature protection

MAX

Range: up to 10°C max above nominal
temperature (for nominal temperature
see label)

MIN

MAX

AUTO

˚C

Using the push/turn control select the
symbol

MAX

-Symbol anwählen.

Hold down the SET key and set the
protection temperature using the
push/turn control.

13.2.2 Undertemperature protection

MIN

Range: from 10°C below minimum
temperature of oven to 10°C above
nominal temperature of oven (for
nominal temperature see label).

The low alarm cannot be programmed
above the value set as high alarm.

Where no undertemperature
protection is required, this has to be
set to the lowest temperature.

MIN

MAX

AUTO

°C

Using the push/turn control select the
symbol

MIN

.

Hold down the SET key and set the
protection temperature using the
push/turn control.

Note:

The temperature monitor can be set independently of the operating mode.
During ramp operation the monitor temperature must always be set sufficiently far above the maximum
working temperature.