Memmert IFP 400 - 800 Incubators (Generation 2003) User Manual
Page 22

PERFECT
page 22
14. Set closure command for second ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g end ) appears.
Hold down the SET key and set next with the push/turn control.
15. Select duration of third ramp segment
Using the push/turn control select the time indication
Hold down the SET key and set the time 08:00 using the push/turn
control.
h
16. Select temperature of third ramp segment
Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of
60 °C using the push/turn control.
˚C
17. Select fan speed for third ramp segment
Turn the push/turn control clockwise until the fan symbol is flashing.
Hold down the SET key and set the fan speed to 50% using the push/
turn control.
18. Set air valve for third ramp segment
Turn the push/turn control clockwise until the air valve symbol is flashing.
Hold down the SET key and set the air valve to 20% using the push/turn
control
19. Set closure command for third ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g end ) appears.
Press the SET key briefly to enter.
20. Exit programme writing mode EDIT
Turn the push/turn control clockwise until EXIT appears on the display.
Press the SET key briefly to enter.
21. Set temperature monitor
Turn the push/turn control clockwise and set the temperature monitor.
(see Section “Temperature monitor”)
MIN
MAX
AUTO
°C
22. Start programme
Turn the push/turn control anticlockwise until the stop symbol
is
flashing.
Hold down the SET key and select Start
with the push/turn control.