22. Set air valve for fourth ramp segment

Turn the push/turn control clockwise until the air valve symbol is flashing.
Hold down the SET key and set the air valve to 0% using the push/turn
control.

The air valve is closed after
sterilisation.

23. Set closure command for fourth ramp segment

Turn the push/turn control clockwise until a segment closure command
(e.g. end ) appears.
Turn the push/turn control clockwise until End appears on the display
and press the SET key briefly to enter.
If the audible signal at programme end endSOUND has been switched on
in SETUP, the oven signals the end of sterilisation.

When the temperature has
reached 30°C, the display shows

End after 1 minute.
The load can now be removed.
The fan continues to run at
maximum speed.

24. Exit programme writing mode EDIT

Turn the push/turn control clockwise until the display shows EXIT and
press the SET key briefly to enter.

25. Set temperature monitor

Turn the push/turn control clockwise and set the temperature monitor,
e.g. ASF +/-5°C.

(see Section “Temperature monitor”)

MIN

MAX

AUTO

°C

26. Start the programme

Turn the push/turn control anticlockwise until the stop symbol

is

flashing. Hold down the SET key and select start

using the push/turn

control.

21.4 Steriliser cassettes

The cassettes should preferably be so arranged in the steriliser that the hot air flow can pass readily through
the air slots.

The load to be sterilised is placed into the steriliser cassettes wrapped in aluminium foil or in steriliser foil
suitable for hot air

(as in the Table in the Section “Guidelines for sterilisation“)

. The air slots in the cassette

must be open for sterilisation.
A temperature probe to confirm the temperature of the load can be introduced through the opening (2).
After sterilisation has been completed the air slots must be closed by moving the slide knob (1).
The sterilised and packed load can then be stored briefly in the closed cassette.