The manually set monitor temperature and the electronic overtemperature protection are monitored on
PERFECT incubators by an adjustable temperature monitor (TWW) Protection Class 3.1 to DIN 12 880, or
by an adjustable temperature limiter (TWB) Protection Class 2 to DIN 12 880. The choice of temperature
monitor is selected in SETUP

(see the menu item Tolerance margin ASF in Section „Basic oven settings”)

13.2.3 Adjustable temperature monitor (TWW) Protection Class 3.1 to DIN 12 880
If the manually set monitor temperatur

MAX

is exceeded, the TWW takes over the control of the

temperature and starts to control at the monitor temperature. As a warning the alarm symbol

is

flashing.

TB approx. 20°C above Tmax

TWW set manually

setpoint temperature

t

°C

controller fault

emergency operation

13.2.4 Adjustable temperature limiter (TWB) Protection Class 2 to DIN 12 880
If the manually set monitor temperature

MAX

is exceeded, the TWB switches off the heating permanently

and can only be reset by pressing the SET key. As a warning the alarm signal

is flashing.

TB approx. 20°C above Tmax

TWB set manually

setpoint temperature

t

°C

controller fault

heating switched off by TWB