7. Set air valve for first ramp segment
Turn the push/turn control clockwise until the air valve symbol is flashing.
Hold down the SET key and set the air valve to 20% using the push/turn
control.
The air valve can be open for
drying during heating.
8. Set closure command of first ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g.end ) appears.
Hold down the SET key and set SPWT [T] with the push/turn control.
The segment closure command
SPW [T] ensures that the
sterilisation time starts only
when the temperature of
180°C has been reached.
9. Select duration of second ramp segment
Using the push/turn control select the time indication.
Hold down the SET key and set the time 01:30 using the push/turn
control.
h
10. Select temperature of second ramp segment
Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of
180 °C using the push/turn control.
˚C
11. Select fan speed for second ramp segment
Turn the push/turn control clockwise until the fan symbol is flashing.
Hold down the SET key and set the fan speed to 100% using the push/
turn control.
The fan is running at maximum
speed.
12. Set air valve for second ramp segment
Turn the push/turn control clockwise until the air valve symbol is flashing.
Hold down the SET key and close the air valve using the push/turn
control.
The air valve is closed during
sterilisation.
13. Set closure command for second ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g. end ) appears.
Hold down the SET key and set next with the push/turn control.