7. Set air valve for first ramp segment

Turn the push/turn control clockwise until the air valve symbol is flashing.
Hold down the SET key and set the air valve to 20% using the push/turn
control.

The air valve can be open for
drying during heating.

8. Set closure command of first ramp segment

Turn the push/turn control clockwise until a segment closure command
(e.g.end ) appears.
Hold down the SET key and set SPWT [T] with the push/turn control.

The segment closure command

SPW [T] ensures that the
sterilisation time starts only
when the temperature of
180°C has been reached.

9. Select duration of second ramp segment

Using the push/turn control select the time indication.
Hold down the SET key and set the time 01:30 using the push/turn
control.

h

10. Select temperature of second ramp segment

Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of

180 °C using the push/turn control.

˚C

11. Select fan speed for second ramp segment

Turn the push/turn control clockwise until the fan symbol is flashing.
Hold down the SET key and set the fan speed to 100% using the push/
turn control.

The fan is running at maximum
speed.

12. Set air valve for second ramp segment

Turn the push/turn control clockwise until the air valve symbol is flashing.
Hold down the SET key and close the air valve using the push/turn
control.

The air valve is closed during
sterilisation.

13. Set closure command for second ramp segment

Turn the push/turn control clockwise until a segment closure command
(e.g. end ) appears.
Hold down the SET key and set next with the push/turn control.