Muster, Ramp – Memmert IFP 400 - 800 Incubators (Generation 2003) User Manual
Page 19

PERFECT
Muster
page 19
10.1 Closure commands for ramp segments
NEXT
Follow-on with next programme segment.
SPWT (T)
SET-POINT WAIT
Wait until the setpoint temperature is reached.
The oven only starts the next programme segment when the programmed
setpoint temperature has been reached, even if the programmed heating
time has already elapsed.
LOOP
Ramp repeat function.
The set programme is repeated after passing through all programmed
segments.
1-99 = repeats
cont = continuous repeat function
HOLD
End of programme without switching off the heating; temperature and
all other settings (e.g. air valve) are maintained.
END
End of programme, heating is switched off, all other settings (e.g. air
valve) are reset to base status.
The programme segments are linked together by the segment closure command. These commands
therefore control the programme sequence.
°C
Ramp
closure
command
ramp
segment 1
spwt (t)
t=time
segment 1
closure
command
ramp
segment 2
next
command
ramp
segment 4
next
closure
command
ramp
segment 5
end
command
ramp
segment 3
spwt (t)
segment 3
segment 1
segment 5
segment 4
delayed
programme start