Muster – Memmert IFP 400 - 800 Incubators (Generation 2003) User Manual
Page 21
![background image](/manuals/567663/21/background.png)
PERFECT
Muster
page 21
5. Select duration of first ramp segment
Turn the push/turn control further clockwise until the time of the first
ramp segment is flashing.
Hold down the SET key and set the time 00:01 using the push/turn
control.
h
6. Select temperature of first ramp segment
Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of
180 °C using the push/turn control.
˚C
7. Select fan speed for first ramp segment
Turn the push/turn control clockwise until the fan symbol is flashing.
Hold down the SET key and set the fan speed to 50% using the push/
turn control.
8. Set air valve for first ramp segment
Turn the push/turn control clockwise until the air valve symbol is flashing.
Hold down the SET key and set the air valve to 20% using the push/turn
control.
9. Set closure command of first ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g. end ) appears.
Hold down the SET key and set SPWT [T] with the push/turn control.
10. Select duration of second ramp segment
Using the push/turn control select the time indication.
Hold down the SET key and set the time 00:45 using the push/turn
control.
h
11. Select temperature of second ramp segment
Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of
180 °C using the push/turn control.
˚C
12. Select fan speed for second ramp segment
Turn the push/turn control clockwise until the fan symbol is flashing.
Hold down the SET key and set the fan speed to 50% using the push/
turn control.
13. Set air valve for second ramp segment
Turn the push/turn control clockwise until the air valve symbol is flashing.
Hold down the SET key and set the air valve to 50% using the push/turn
control.