16 Heating power distribution BALANCE

On ovens series INP 500-800 and UNP/UFP 400-800 it is possible in SETUP to correct the heating power
distribution BALANCE between the top and bottom heater groups to suit the individual application. The
adjustment range is –50% to +50%.
The setting 0% restores the factory-set heating power distribution.

top

heater power

top

heater power

bottom heater

power

-20%

bottom heater

power

+30%

Example 1:

With the setting –20% the bottom heater
groups heat with 20% less power than the top
heater groups.

Example 2:

With the setting +30% the bottom heater
groups heat with 30% more power than the
top heater groups.