Maintenance and Servicing

Fig. 36

Dismantling the right and left inner walls.

Remove the silicone plugs and 8 screws on

each side and pull out the wall.

A

Fig. 37

Pull the temperature sensors (A) out of the

holders and unscrew the test chamber rear

wall

8.3

Regular maintenance

Every three months if in permanent operation

Grease the moving parts of the doors (hinges

and lock) with thin silicone grease and check

that the hinge screws are not loose.

Check that door seals fit tightly, adjust door

if necessary (see Chapter 8.4 ).

Check that the condensation water runoff

tube ( Fig. 38 ) at the rear of the chamber

interior is not blocked and clean if necessary.

Every six months

Check air filter of cooling unit for dirt and

clean if necessary (see Chapter 8.5 ).

Annually:

Clean air filter of cooling unit (see Chapter 8.5 ).

Grease the moving parts of the doors (hinges and lock) with thin silicone grease and check

that the hinge screws are not loose.

Check that door seals fit tightly, adjust door if necessary (see Chapter 8.4 ).

Check the safety labels and renew warning stickers if necessary (see Fig. 1 on page 7 ).

Fig. 38

Condensation water runoff

tube