Advanced functions
The ASF alarm goes off automatically as soon as 50 % of the set tolerance band of the set-
point (in the example 50 °C ± 1 °C) is reached again (section A).
If the temperature setpoint is altered, the ASF is automatically disabled temporarily (see in
the example: The setpoint is changed from 50 °C to 25 °C, section D), until it has reached the
tolerance range of the new temperature setpoint (section E).
52°C
48°C
ASF active
52°C
48°C
ASF alarm
ASF active
ASF active
27°C
23°C
°C
50°C
25°C
t
A
B
C
D
E
AUTO
AUTO
AUTO
Fig. 29
Schematic diagram of how the ASF temperature monitoring functions
Switching on the automatic temperature monitor:
Select the AUTO icon with the push-turn control.
Hold down the SET key and set to on with the push-turn control.
MIN
MAX
AUTO
°C
Switching off the automatic temperature monitor:
Select the AUTO icon with the push-turn control.
Hold down the SET key and set to off with the push-turn control.
MIN
MAX
AUTO
°C
The tolerance band for the ASF can be set in the SETUP in the range 2 ... 20 °C (see Chap-
ter 7.2 Basic appliance settings (Setup) on page 46 ).