50

Advanced functions

The ASF alarm goes off automatically as soon as 50 % of the set tolerance band of the set-

point (in the example 50 °C ± 1 °C) is reached again (section A).
If the temperature setpoint is altered, the ASF is automatically disabled temporarily (see in

the example: The setpoint is changed from 50 °C to 25 °C, section D), until it has reached the

tolerance range of the new temperature setpoint (section E).

52°C

48°C

ASF active

52°C

48°C

ASF alarm

ASF active

ASF active

27°C

23°C

°C

50°C

25°C

t

A

B

C

D

E

AUTO

AUTO

AUTO

Fig. 29

Schematic diagram of how the ASF temperature monitoring functions

Switching on the automatic temperature monitor:
Select the AUTO icon with the push-turn control.
Hold down the SET key and set to on with the push-turn control.

MIN

MAX

AUTO

°C

Switching off the automatic temperature monitor:
Select the AUTO icon with the push-turn control.
Hold down the SET key and set to off with the push-turn control.

MIN

MAX

AUTO

°C

The tolerance band for the ASF can be set in the SETUP in the range 2 ... 20 °C (see Chap-

ter 7.2 Basic appliance settings (Setup) on page 46 ).