Putting into Operation

4.5

Connecting

Caution:

Observe the country-specific regulations when making connections (e.g. in Germany

DIN VDE 0100 with an RCD circuit breaker). Also observe the connected loads and

power values (see chapter 2.9 Technical data on page 17 ).

The test chamber is intended for operation on an electrical power system with a system im-

pedance Z

max

at the point of transfer (service line) of a maximum of 0.292 Ohm. The operator

must ensure that the test chamber is operated only on an electrical power system that meets

these requirements. If necessary, you can ask your local energy supply company what the

system impedance is.
Only appliances may be connected to external connections whose interfaces comply with the

requirements for safety extra-low voltage (e.g. computer, printer).

Switch on appliance by pressing the main switch on the front of the appliance (see Fig. 18

on page 27 ).

Caution:

The first time the appliance is operated, it must not be left unattended until it has

reached the steady state.