Operation and control

15. Setting the duration of the third ramp segment:

Select the time display with the push-turn control.

Hold down the SET key and set the time to 01:00:00 AM

with the push-turn control.

h

16. Setting the temperature of the third ramp segment:

Turn the push-turn control to the right until the tem-

perature display flashes.

Hold down the SET key and set 37.0 °C with the push-

turn control.

°C

17. Setting the fan speed of the third ramp segment:

Turn the push-turn control to the right until the fan icon

flashes.

Hold down the SET key and set the fan speed of 40 % (4

bars) with the push-turn control.

18. Setting the relative humidity of the third ramp segment

(only for CTC 256):

Turn the push-turn control to the right until the humidity

display flashes.

Hold down the SET key and set the desired humidity

setpoint of 50.0 % rh with the push-turn control.

%

rh

19. Setting the close command of the third ramp segment:

Turn push-turn control to the right until a close command

appears in the display, e.g. end, and press the SET key briefly

to confirm.

20. Leave programming write mode EDIT:

Turn push-turn control to the right until EXIT appears in

the display, and press the SET key briefly to confirm.

21. Setting the temperature monitoring:

Turn the push-turn control to the right and adjust the

temperature monitoring (for more detailed information

on this, see page 48 ).

MIN

AUTO

MAX

°C

22. Starting the programme:

Turn the push-turn control to the right until the stop

icon flashes.

Hold down the SET key and select the start icon with

the push-turn control. Release the SET key, and the pro-

gramme is started.