Memmert TTC 256 Temperature test chamber (Generation 2003) User Manual

Page 47

background image

47

Advanced functions

Date
The controller contains a calendar which automatically

accounts for the different lengths of months, and for leap

years.

Weekday

Tu

Year
Adjustment range: From 2000 to 2100

Acoustic signal at programme end
ENDSOUND
Setting: OFF oder ON

Acoustic signal for alarm, e.g. over/undertemperature
ALARM sound
Setting: OFF or ON

Communication address
Adjustment range: 0 bis 15 (see Chapter 7.6 Communication

interfaces on page 55 )
Tolerance band ASF
Adjustment range: 2 to 20 °C (see page 49 )

Language
Setting: GERMAN, English, franCAIS, ESPANOL und italIANO

Compensation correction values (CAL 1-3, rh20, RH90) for cali-

bration of temperature and humidity at customer site
Chapter 7.4 Calibration on page 51 )

Automatic defrosting system

(for details on settings, see page 54 )

Leave Setup
Save all settings and leave the SETUP mode
The real time clock which is set in the SETUP contains the date and clock time. The real time

clock is used for logging purposes in accordance with GLP. Date and clock time are specified

on the log printout. On graphical printouts, the time axis is labelled with the real time. The

clock is battery-buffered and independent of the mains connection. The integrated lithium

battery of the type CR 2032 has a lifetime of approx. 10 years.