beautypg.com

2 humidity calibration (only for ctc 256) – Memmert TTC 256 Temperature test chamber (Generation 2003) User Manual

Page 53

background image

53

Advanced functions

4. The reference instrument should now display 30 °C after the calibration procedure.
5. With CAL.1, another compensation temperature below CAL.2 can be programmed in the

same way, and with CAL.3, one above it.

If all compensation correction values are set to 0.0 °C, the factory calibration settings are

restored.

7.4.2 Humidity calibration (only for CTC 256)

The test chamber can be calibrated for the individual customer by means of two balance

points at 20 and at 90 % relative humidity. For each selected balance point, a positive or nega-

tive compensation correction value can be set between and –10 % and +10 %.

rh 20

+3,0 %

rh 90

-2,0 %

Calibration at factory

60%

40%

0%

20%

80%

100%

Fig. 31

Humidity calibration (example)

Setting:
1. Set the desired humidity balance point in the SETUP (see Chapter 7.2 Basic appliance set-

tings (Setup) on page 46 ) and set the corresponding compensation correction value to 0.0

°C.

2. With a reference instrument, measure the deviation in the stationary state in the selected

humidity balance point.

3. Setting the compensation correction values in the SETUP. If the measured reference humid-

ity is too low, the compensation correction value must be set with a negative sign.

4. Perform a control measurement with the reference instrument.
The procedure can be performed with humidity balance points of 20 % rh and 90 % rh.
Example:
Humidity deviation at 90 % should be corrected.
1. Set humidity balance point in the SETUP to RH 90 and set the accompanying compensation

correction value to 0.0 %rh:

PRINT

SETUP

loop

t3

t4

t2

t1

on

off

Mo

Tu

We

Th

Fr

Sa

Su

3

4

2

1

STERI

DEFRO

°C

°C

rh

mb

%

CO

mb

2

IN 1

IN 2

OUT

MIN

AUTO

MAX

rh

b

%

%

%

This manual is related to the following products: