Operation and control

PRINT

SETUP

Normal mode

(see page 30 )

Week time

switch

(see page 33 )

Programme

mode

(see page 35 )

Printer

(see page 46 )

Basic appliance

settings

(see page 46 )

Fig. 22

Operating modes

5.8

Operating mode - settings

1. Switch on appliance by pressing the main switch (main switch comes out of

appliance, see Fig. 19 ).

2. Hold SET key down for approx. three seconds, the selected operating mode

then begins to flash.

3. Select the desired operating mode (normal mode, week time switch, pro-

gramming mode, printer or basic appliance settings) by turning the control

with SET key held down.

4. Release the SET key, and the selected operating mode is saved.

5.8.1 Normal mode

The test chamber runs in permanent operation in this operating mode, at the values set on

the operating panel (see above). Example of settings: see next Chapter 5.8.2 .
1. Load test chamber (see page 26 ).
2. Switch on appliance by pressing the push-turn control on the operating panel so that it

comes out of the appliance ( Fig. 19 ).

3. Select the normal operating mode

with the push-turn control:

PRINT

SETUP

4. As described above, set the individual parameters with the push-turn control and the

SET-key:

Temperature setpoint

Adjustment range: - 42°C to 190°C

°C

Fan speed

Adjustment range: 10 % to 100 % in 10-%

steps

At low temperatures the fan speed is automatically reduced. In strong heating operation

it is automatically increased.

set

set

set

set