beautypg.com

Memmert TTC 256 Temperature test chamber (Generation 2003) User Manual

Page 38

background image

38

Operation and control

SET-POINT WAIT (H – humidity, only for CTC 256)

Wait until setpoint humidity has been reached.

Appliance starts the next programme segment only when the

programmed setpoint humidity has been reached, even if the set

heating up time has already elapsed.

SET-POINT WAIT (TH – temperature and humidity, only for CTC

256)

Wait until setpoint temperature and setpoint humidity have been

reached.

Appliance starts the next programme segment only when the

programmed setpoint temperature and programmed setpoint

humidity have been reached, even if the set heating up time has

already elapsed.
Ramp repeat function

The programme entered is repeated after it has run through all

programmed segments.

1-99 = repeats

cont = endless repeat function
Programme end while maintaining the temperature and humidity

of last programme ramp
Programme end, switching off the heating /cooling function and

humidity

Close

command

ramp

segment

No. 5

end

Segment5

°C

t=time

Delayed

programme start Segment1

Close

command

ramp

segment

No. 1

spwt (t)

Segment2

Close

command

ramp

segment

No. 2

next

Segment3

Close

command

ramp

segment

No. 3

spwt (tH)

Segment4

Close

command

ramp

segment

No. 4

next

Fig. 25

Schematic example of the use of ramp segment close commands