57

Advanced functions

7.6.4 Log memory

The controller continually logs all relevant measured values, settings and error messages at

1-minute intervals. The internal log memory is listed as a ring memory, i.e. the oldest log data

are always overwritten automatically with new data.
The logging function cannot be switched off, but is always active. The measured data are

stored in the controller, safe from manipulation. For documentation purposes the controller

memory can be read out via the Celsius control software. Each dataset is stored with a unique

timestamp.
The internal log memory has a size of 1024 kB. This corresponds to a storage capacity for

about three months in permanent operation.
Since large amounts of data are stored during ramp mode, this can reduce the maximum log-

ging time.
If the power supply is interrupted, the time of the power cut and the return of voltage are

stored in the controller.

Reading in the log memory to the PC via USB interface

The log data can be read out either via the USB interface and then printed out from the

PC or via a connected PCL3-compatible printer.
Using the “Celsius” programme the log memory of the controller can be read out to a PC

and from there be displayed graphically, printed out and stored.

The log memory of the controller is not modified or deleted by the reading out.

Printing out log memory

(see also Chapter Printer on page 46 )
For a printout, the GLP header is also printed automatically, and contains the following details:

Date of printout

Period of logging

Consecutive page numbers

Serial numbers and appliance name

If the printer is not ready (e.g. ink cartridge or paper tray empty), no log data will be lost.

Multiple printouts can also be made, since the log memory is not deleted after printing.