beautypg.com

4 calibration – Memmert TTC 256 Temperature test chamber (Generation 2003) User Manual

Page 51

background image

51

Advanced functions

7.3.3 Warning messages

A repeated acoustic signal indicates an error in the temperature control system or in the hu-

midification system. It is set off in the following cases:

Overtemperature limit is triggered. Check the MAX setting

and increase if necessary (see Chapter 7.3.1 Electronic tem-

perature monitoring (TWW) on page 48 ).
Undertemperature limit is triggered. Check the MIN setting

and reduce if necessary (see Chapter 7.3.1 Electronic tem-

perature monitoring (TWW) on page 48 ).

Automatic monitoring function is triggered. Check the ASF-set

limit value in the SETUP menu (see Chapter 7.2 Basic appli-

ance settings (Setup) on page 46 ).

Water tanks are empty. Fill up with water (see page 22 ).

Humidity exceeds the preset setpoint for longer than 30

minutes. Open door for 30 sec. and wait to see if the control-

ler steadily adjusts to the setpoint. If the error occurs again,

contact Customer Service.

This monitoring function only starts to work once the humidity setpoint has been

reached.
The acoustic alarm can be temporarily switched off by pressing the SET key.

7.3.4 Safety and monitoring equipment for the cooling unit

A high pressure limiter and a low pressure monitor protect the cooling unit from inadmis-

sible operating pressure. When the high pressure limiter is triggered, the heaters and seals are

switched off, and the caution icon lights up.
The condensation pressure is controlled by the automatic speed adjustment of the condenser

fan. The fan can therefore automatically run faster or slower in operation, depending on the

thermal energy extracted. The condenser fan may continue running for a short time after the

main switch has been turned off, to cool down the condenser.
The condenser is fitted with an overheating limit. If the motor current increases, together with

an increase in the coil temperature, the condenser is switched off. After the coil temperature

has fallen, the overheating limit switches off automatically.

7.4

Calibration

7.4.1

Temperature calibration

The test chamber can be calibrated customer-specifically using three calibration temperatures

of your choice:

CAL.1

Temperature calibration at low temperature

CAL.2

Temperature calibration at medium temperature

CAL.3

Temperature calibration at high temperature

For each selected balance point ( Fig. 30 ), a positive or negative compensation correction value

can be set between –4.9 ºC and +4.9 ºC. There must be a difference of at least 10 ºC between

each of the individual balance points.

This manual is related to the following products: