3 preventative maintenance schedule, Preventative maintenance schedule, Cleaning and maintenance – Leica Biosystems ST5020 User Manual
Page 77
![background image](/manuals/565513/77/background.png)
77
Leica ST5020
6.3 Preventative maintenance schedule
Maintenance activity
Clean the TFT screen with a lint-free cloth.
Follow the manufacturer's instructions when using
screen cleaner.
Clean outside surfaces of instrument and
lid with soft cloth.
Use cleaning detergent, if necessary.
Clean wash vessels.
Check the wash vessel O-rings for damage and lubrica-
te them slightly.
See
Fig. 4, page 16
for further details.
Remove and clean the reagent vessels. A dishwasher
up to 65 °C can be used.
Check waste water hose for accumulated dirt and cle-
an, if necessary.
Remove paraffin residues from paraffin collecting
tray(s) of the oven(s).
Check strainer (V-Filter, Pos. 3 in
Fig. 6, S. 18
) in tap wa-
ter inlet hose for accumulated dir tand clean, if neces-
sary.
Check air vent holes of fans for dust. Clean, if necessa-
ry.
Check the tap water inlet hose connection for damage.
If necessary, replace the gasket as described in
Chap-
ter 4.3, page 18
.
Replace activated carbon filter as described in
Chapter
4.6, page 24
.
Check electrical connections at rear panel of
instrument.
daily
√
√
weekly
√
√
√
√
monthly
√
√
√
√
√
6.
Cleaning and maintenance