Operation – Leica Biosystems ST5020 User Manual
Page 47

47
Leica ST5020
Fig. 44
Deleting a reagent
•
Highlight the reagent you want to delete from the list.
•
Press DELETE and confirm the screen prompt. The reagent is removed
from the list and the blue highlighting bar moves one row down.
Reagents can only be deleted if they are NOT allocated to any
program or station. Otherwise there will be a message that de-
leting is not possible.
If the Reagent Management System
(RMS) is switched on, the table head-
ers are displayed as buttons.
The following limit parameters can be defined
for the RMS:
•
Maximum number of days the reagent may
be used.
•
Maximum number of racks that can be
processed until the next reagent change.
In ‘instrument status’ a red bar in the
station button indicates how long a
reagent has already been used de-
pending on the values entered here.
The longer the red bar, the closer the
time when the reagent will have to
be changed. (See chapter 5.5.1)
Press the header buttons to access the corre-
sponding numerical keypads for entering RMS
data.
Numbers between 1 and 99 can be entered in
both columns.
↓
↓
↓
Setting RMS data
5.